Our Solutions

Flow Calibrator CF-1: An Innovation in Field Calibration

CF-1 flow calibratorThe Flow Calibrator model CF-1 is a robust and compact calibrator designed for field use. At the heart of this flow calibrator is a laminar flow element. In this component, the medium flows through a bundle of small tubes, creating a laminar flow. Due to the resistance in this element, a pressure difference can be measured. This pressure difference is directly proportional to the flow rate.

Wide Applicability and High Precision

The dynamic range of the CF-1 is 1:100. For applications such as personal sampling, the unit is equipped with a measurement cell of 50 – 5000 Ncc/min. For high flow applications, such as particulate matter samplers, it features a measurement cell of 0.45 – 45 Nl/min. The unit is extremely compact and can be connected on both the suction and blow sides. Ambient pressure and the temperature of the medium (air) are measured during flow measurement. Reporting is available in both normalized and actual conditions. Additionally, the data is integrated over time, allowing for volume logging. Optionally, the unit can be supplied with an external temperature and relative humidity sensor.

Measurable Parameters and Software

With the CF-1, the following parameters can be easily determined:

  • Flow (actual and normalized)
  • Volume (actual and normalized)
  • Atmospheric pressure
  • Temperature and relative humidity (with external sensor)

The standard supplied software allows you to apply correction factors after re-certification of the instrument for the most accurate readings.

Continuous Monitoring and Data Logging

Another application enables continuous logging of all measured and calculated parameters to your PC over time.

Technical Specifications

Range:0.45 – 45 Nl/min
50 – 5000 Ncc/min
Connections:¼” gas thread
Accuracy:Standard 1.5 %F.S.
High Accuracy 0.8 %R + 0.1 %F.S.
Repeatability:0.5 %F.S.
Response Time:500 ms
Environmental Conditions:0 – 1050 mbar
0 – 40 °C
Dimensions:145 x 73 x 58 mm
Weight:300 grams
Battery:LiPo 4000mA
Housing:ABS with rubber protection
Sensors:
Barometric Pressure:
Range:10 – 1050 mbar
Hysteresis and Linearity:0.25 %F.S.
Resolution:0.1 mbar
Accuracy:1 %R + 2.5 mbar
Temperature:
Range:-10 – 60 °C
Resolution:0.1 °C
Accuracy:1 %R + 0.4 °C
Optional Sensor:
Cable Length:1 meter
Range:-10 – 60 °C
0 – 100 %RH
Resolution:0.1 °C
0.5 %RH
Accuracy:1 %R + 0.4 °C
+2 %RH between 10 and 90 %RH at 25 °C

Get in touch with one of our specialists

Ravebo stands for sustainable cooperation with its clients. Both within Ravebo and with our manufacturers, we work with short and direct lines — and that’s practical! We are happy to get started with your request for your application in measuring and monitoring dust and gases.